Applanation tonometers

We repair and calibrate
applanation tonometers

01.
Cleaning and disinfection

Upon arrival, the tonometer is cleaned and disinfected

02.
Diagnosis

Disassembly and check the condition of all parts

03.
Estimate

An estimate is drawn up within 24 hours of receipt

04.
Repair

Repair performed within 24 hours of receipt of acceptance 

01
Review of the internal mechanism

Grinding of the main shaft is performed with a precision lathe

02
Replacement of damaged parts

Where excessively damaged parts are replaced

03
Supply of courtesy tonometer

If requested, a courtesy tonometer will be provided while yours is being calibrated

Does my tonometer really need an overhaul?

How to Test if
the tonometer is reliable

Fit the prism in the cone holder, rest the tonometer on the slit lamp support, check that by turning the graduated ring nut the rod oscillates around the zero value.

Stringere la luce di fessura e proiettarla sull'apice del prisma. Controllare che arrivi sempre, sia con il cono arretrato che in avanti

Verificare l'oscillazione con l'asta di prova i valori di 0 mmHg 20(2) mmHg e  60(6) mmHg. La tacca centrale corrisponde a 0, la prima tacca a 2 e la terza a 6. La più importante come evidenziato in basso è la posizione 2

The importance of position "2"